Walchem WebMaster Instruction Manual Page 39

  • Download
  • Add to my manuals
  • Print
  • Page
    / 129
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 38
35
sample stream to resume control. If failed, refer to Section 7 for troubleshooting help. A
calibration will fail if the offset calculated is more than 200 mV, or if the slope calculated is
outside of 0.5-1.5.
1 pt Buffer Cal
Step 1 will explain that control is suspended during the calibration procedure. Press the Enter key
when the cursor is on “Continue” to start the calibration. To cancel the calibration, press the Next
key to move the cursor to “Cancel”, then press the Enter key.
Step 2 will prompt you to rinse the electrode, and place it into the buffer solution. Press the Enter
key to continue.
Step 3 will ask you to enter the value of the buffer solution. Use the Arrow keys to change the
value, then press the Enter key to highlight ‘CONTINUE’, then press Enter again to put the value
into memory.
Step 4 will display the mV output of the electrode in the buffer. When this reading is stable, press
the Enter key to continue.
At the end of the calibration, the display will tell you if the calibration was successful or if the
calibration failed. If successful, press the Enter key after the electrode has been replaced into the
sample stream to resume control. If failed, refer to Section 7 for troubleshooting help. A
calibration will fail if the offset calculated is more than 200 mV, or if the slope calculated is
outside of 0.5-1.5.
1 pt Process Cal
Step 1 will display the current pH reading of the system water. Below that, use the Arrow keys to
fill in the new value, then press the Enter key to highlight ‘CONTINUE’, then press Enter again to
put the value into memory.
At the end of the calibration, the display will tell you if the calibration was successful or if the
calibration failed. If successful, press the Enter key after the electrode has been replaced into the
sample stream to resume control. If failed, refer to Section 7 for troubleshooting help. A failed
calibration is caused by an offset from the uncalibrated reading of more than 200 mV.
4-20 mA Input 1-8 Calibration
Calibrate 4 mA
Following the instructions for your 4-20 mA transmitter, supply the WebMaster controller with a 4
mA signal from the transmitter. You should verify that the transmitter is accurate by measuring the
output with an ammeter. You can read the present value of the mA signal as measured by the
controller on the top line of the display. If the value is not exactly 4 mA, press the Enter key when
the cursor is on “Calibrate 4 mA”. Once this is done, the controller will read the transmitter signal
as exactly 4 mA.
The transmitter signal must be +/- 2 mA of the expected 4 mA value, or else the calibration will
fail.
Calibrate 20 mA
Repeat this process, supplying the controller with a 20 mA signal from the transmitter, and press
the Enter key when the cursor is on the “Calibrate 20 mA” to make the present value exactly 20
mA.
The transmitter signal must be +/- 2 mA of the expected 20 mA value, or else the calibration will
fail.
Page view 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 128 129

Comments to this Manuals

No comments